Beckman Coulter Optima L-90K Ultracentrifuge

Beckman Coulter Optima L-90K Ultracentrifuge

The Beckman Optima L-90K Ultracentrifuge offers the fastest separations possible in the minimum amount of time with its performance capability of 694,00 x g at up to 90,000 rpm. This versatile floor model is compatible with a wide range of superior rotors including zonal and continuous flow for large-volume separations. The Optima L-90K is perfect for applications such as subcellular particles and virus isolations, and high-purity plasmid DNA.

Available rotors:

  • 50.2 Ti (12 x 39 ml)
  • 70.1 Ti (12 x 13.5 ml)
  • SW28 Ti (6 x 38.5 ml)
  • SW40 Ti (6 x 14 ml)

 

*The Basic Lab Support Section team does not provide centrifuge tubes and adapters. Please purchase and use the tubes and adaptors needed for your experiment.*

For Ultracentrifuge

Use only specified tubes and adaptors by the manufacturer

For High-Speed centrifuge

Disposable tubes and bottles can be used.
Make sure to use matched sets of tubes, bottles, and adaptors.

 

 

Share on:

Features

Featuring an imbalance-tolerant drive, the L-90 K provides unmatched reliability with its highly efficient vacuum-encased induction drive which allows smooth and quiet performance. Plus, this ultracentrifuge model is sized for efficient operation within laminar flow hoods.

Equipped with easy-to-use, environment-friendly and microprocessor–based control cooling systems, the Beckman L-90 K eliminated the use of CFCs and other dangerous liquid refrigerants. This Beckman centrifuge also has an overspeed system to ensure that rotors do not exceed maximum speed.

Other features of the Beckman Optima™ L-90 K centrifuge include digital displays and control via touch-sensitive keys, choice of slow-start and standard acceleration, built-in w2t integrator, air-cooling drives, and CFC-free thermoelectric cooling and heating modules.

 

Specifications:

  • Max speed: 90,000 rpm.
  • Speed control: +/- 20 rpm of set speed.
  • Acceleration: maximum acceleration from 0 rpm to set speed, or slow acceleration from 0 to 500 rpm, followed by maximum acceleration to set speed.
  • Deceleration: full dynamic braking to 0 rpm, slow (full brake to 500 rpm followed by reduced braking to a gentle stop), or no brake.
  • Imbalance tolerance: +/- 5 mL or 10% whichever is greater.
  • Set time: up to 99 hours and 59 minutes.
  • Set temperature: 0 to 40 °C in increments of 1 °C.
  • Temperature control: +/-0.5 °C of set temperature.
  • Ambient temperature range: 15 to 40 °C.
  • Noise level measured 0.91 m (3 ft) in front of the centrifuge: 57 dBa.